28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 6.2 Mixing Head Position Adjustment
3 - 85
3
Search
Book TOC
Go Back
VP - 6.2 Mixing Head Position Adjustment
Purpose:
Perform this procedure to ali
g
n the Mixin
g
Assembly.
1. Perform Check procedure “VP - 6.1 Mixin
g
Head Position Check” (Pa
g
e 3-77).
2. Perform VP-3.8,VP-3.10, VP-3.16 checks.
3. Perform vertical adjustment.
a. Loosen top Ball Slide Screw (Fi
g
ure 3-41).
b. Adjust Mixin
g
Assembly until Paddle is parallel to strai
g
ht ed
g
e (Fi
g
ure 3-42, Pa
g
e-3-87).
c. Ti
g
hten Ball Slide Screws.
4. Adjust Head Position.
E and F Loader
a. Loosen Mixin
g
Assembly screws (Fi
g
ure 3-43, Pa
g
e-3-88).
b. Adjust Mixin
g
Assembly until “VP - 6.1 Mixin
g
Head Position Check” (Pa
g
e 3-77) is OK.
c. If E Loader, adjust position at both paddles.
Special Tools
DESCRIPTION QTY PN
Mixin
g
Head
Adjustment Tool
1 9420082
Rack Guide
Ali
g
nment Tool
1 9420084