3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING
IP - 1 Blood Detected In Sample Line, 2 of 3
1 - 11
1
Search
Book TOC
Go Back
IP - 1 Blood Detected In Sample Line, 2 of 3
Problem
Corrected
?
Perform Total
Call
Yes
1
- Turn OFF
Sample Loader
- Reseat CN5 on
CPU Main Bd
- Reseat
Connector on
Blood In-Line
Sensor Bd
- Perform Blood
In-Line Sensor
Check (VP-12.1)
Replace Sample
Line (PL-7.8)
Problem in
Line
?
Pull Sample Line Through
Sensor About 1/2 "
(1.27cm), and Inspect Line
For Dark Spots and Kinks
That Might Cause Error
Yes
No
No
2
Blood
In-Line
Sensor
SL-267