EasyManua.ls Logo

Abbott CELL-DYN 3000 - Appendix 3 A - Sensors, LEDs and POTs

Abbott CELL-DYN 3000
296 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
Appendix 3 A - Sensors, LEDs and POTs
3 - 168
3
Search
Book TOC
Go Back
Appendix 3 A - Sensors, LEDs and POTs
Table 3 - 7
Sensor LED Test Point Normal Voltages POT VP#
A Rack Sensor Alarm LED I CE MARK: TP 5
NON-CE MARK: TP
Header, Pin 16
CE MARK Blocked: 0.5 -
1.0V
NON-CE MARK
Blocked: 5V
UnBlocked: 0v
R5
VR1
4.1
4.2
B Rack Sensor
(above E
Sensor)
Alarm LED II CE MARK: TP 4
NON-CE MARK: TP
Header, Pin 14
CE MARK Blocked: 0.5 -
1.0V
NON-CE MARK
Blocked: 5V
UnBlocked: 0V
R4
VR2
4.1
4.2
C Rack Sensor Alarm LED III CE MARK: TP 3
NON-CE MARK: TP
Header, Pin 12
CE MARK Blocked: 0.5 -
1.0V
NON-CE MARK
Blocked: 5V
UnBlocked: 0V
R3
VR3
4.1
4.2
D Rack Sensor Alarm LED IV CE MARK: TP 2
NON-CE MARK: TP
Header, Pin 10
CE MARK Blocked: 0.5 -
1.0V
NON-CE MARK
Blocked: 5V
UnBlocked: 0V
R2
VR4
4.1
4.2

Table of Contents