TROUBLESHOOTING
IP - 2 Erratic Results, 2 of 2
1 - 14
1
Search
Book TOC
Go Back
IP - 2 Erratic Results, 2 of 2
Result
Precision
OK
?
Perform Total Call
1
Yes
Inspect/Correct Wash Block(s)
- Ensure Wash Tubing Connections are
OK
- Perform Sample Needle Check VP-9.1
- F Loader- Perform Vent Needle Check
VP-9.2
- G Loader- Check Operation of Lower
Wash Solenoid (Ensure Tubing is OK)
Set
BAR (1/0) Switch
to
1
Press
MAN
key
- Perform Lead Bubble Check VP-13.1
- Run Sample 10 Times
No
Inspect/Correct Mixing Motor(s)
- Perform All Mixing Checks, VP-6
- 3500 (E Loader) Only- Ensure Mixing
Motors are Connected Properly:
- Set
BAR (1/0) Switch
to
0
- Press
R
Mixing Motor 1 (Left) Should Rotate
- Set
BAR (1/0) Switch
to
1
- Press
R
Both Mixing Motors Should Rotate
If error persists, See CELL-DYN 3000/
3500 System Troubleshooting Guide
W OC Errors IP-9
HGB Errors IP-27
RBC/PLT/MCV/W IC Errors IP-29