TROUBLESHOOTING
IP - 22 Alarm 2, Mixing Failure 4 of 6
1 - 85
1
Search
Book TOC
Go Back
IP - 22 Alarm 2, Mixing Failure 4 of 6
Problem
Corrected
?
Perform Total Call
No
B1
- Turn Sample Loader Power OFF
- Check Contact Sensor Wiring
[CE MARK]
Reseat Contact Sensor In-Line
Connector (on Tower)
[ALL]
Verify wires to Contact Sensor Bd. are
not broken
- Verify Good Electrical Connection at
CPU Main Bd. CN5
- Clean Sensor
- Push Up Paddle(s) with Finger several
times, movement should be smooth, not
sticking.
B
SL-121E
Mixing
Contact
Sensor
Mixing
Home
Sensor
Mixing Limit
Sensor
[CE MARK]
Contact Sensor
In-Line Connector
Yes