CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 8.1 Needle Sensors Check/Adjustment
3 - 121
3
Search
Book TOC
Go Back
VP - 8.1 Needle Sensors Check/Adjustment
Purpose:
Perform this procedure to verify and adjust Aspiration Needle Home and Limit Position.
1. Remove Racks and any loose items under Needle(s).
2. Put Sample Loader in Manual Mode.
a. Press and Hold
MAN
and
START
.
b. Turn Power
ON
, hold keys down until BEEPING stops.
3. Set
BAR (1/0) Switch
to
1
.
4. Verify Aspiration Needle is free in Wash Block.
5. Ensure Needle(s) are pushed all the way DOWN to the collar rin
g
and clamp is ti
g
ht (Fi
g
ure
3-64).
6. Ensure the hole in needle is facin
g
in the proper direction (Fi
g
ure 3-65, Pa
g
e-3-123).
Special Tools
DESCRIPTION QTY PN
None