TROUBLESHOOTING
IP - 1 Blood Detected In Sample Line, 1 of 3
1 - 10
1
Search
Book TOC
Go Back
IP - 1 Blood Detected In Sample Line, 1 of 3
Blood in
Line
?
Check for Blood in
Sample Line Near
Blood In-Line
Sensor
1
No
START
From the Data Station Main Menu
- Go to Special Protocols and Press
CLEAN NEEDLE Key
- Check All Tubing, T-Connectors, and
Y-Fittings leading Through SOL #96
and #92 Ending at Waste Chamber
#2
- Check 12" (82.7 kPa) Vacuum Used
for Evacuating Leftover Sample
- Check Diluent Line leading Through
SOL #91 along With 4 PSI (27.6 kPa)
Used to Flush the Line
- Check Vacuum Connection from
Waste Chamber #2 to Sample
Loader Wash Block
YES