CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 3.15 Pusher Arm to Platen Adjustment
3 - 53
3
Search
Book TOC
Go Back
VP - 3.15 Pusher Arm to Platen Adjustment
Purpose:
Perform this procedure to ali
n the X-Axis Pusher Arms in the Platen Cutouts.
CAUTION:Do not remove Platen to perform this adjustment.
1. Perform Check procedure “VP - 3.14 Pusher Arm to Platen Check” (Pa
e 3-51).
2. Loosen 2 Pusher Arm mountin
screws (1/2 turn), adjust pusher arm, ti
hten screws (Fi
ure
3-24).
3. Loosen 2 screws at back of pusher head, adjust as required, and ti
hten screws.
4. Go to “VP - 3.16 X-Axis Solenoid to Rack Guide Check” (Pa
e 3-55).
Special Tools
DESCRIPTION QTY PN
None