TROUBLESHOOTING
IP - 6 Sample Loader Dead, No Lights, No sounds, 1 of 4
1 - 22
1
Search
Book TOC
Go Back
IP - 6 Sample Loader Dead, No Lights, No sounds, 1 of 4
In-Line Fuse
SL-113B
Voltage Select
Switch [OLD Only]
Insert Flat Blade
Screwdriver Here
to Remove
Power Input
Panel
START
Yes
Check Fuse (Replace if Necessary;
PL-10.13
)
120VAC: 1.6A FB 250VAC
220/240VAC: 2.0A SB 250VAC
Yes
1
Problem
Corrected
?
No
Yes
- Verify Only Ten (10) Racks
Used
-
[NON-
CE MARK]
Check
Voltage Select Switch
-
[CE MARK]
Check Polarity
of Keyboard Cable (Red
stripe on cable must face
CN2 pin 1 on Keyboard).
- Perform Power Supply
Check
VP-1
All Voltages
OK
?
Perform Total Call
All Voltages
Missing
?
No
No
2