3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING
IP - 17 Rack Fault or Alarm 4 (m, n, o, p, q, r) Rack Sensors, 2 of 3
1 - 68
1
Search
BookTOC
GoBack
IP - 17 Rack Fault or Alarm 4 (m, n, o, p, q, r) Rack Sensors, 2 of 3
Problem
Corrected
?
- Perform VP-3.8 Rack to
Laser Shield Check
- Perform VP-3.16 X-Axis
Solenoid to Rack Guide
Check
- Load 10 Normal Racks in
Normal Mode
- Press RUN
- Allow Racks to Cycle 15
Minutes
Perform Total
Call
Yes
No
1
2
SWITCHES
RACK SENSORS (UNBLOCKED)
(BLOCKED BY
A RACK)
CPU MAIN BD. A B C D E FROM TO
[NON-CE MARK]
TP Pin #161412108
0V
+5.0V
[NON-CE MARK]
CN3 Pin # 5 7 9 11 13
0V
+1to 1.5V
[CE MARK]
TP # 5 4 3 2 1
0V
+0.5 to 1.0V
[CE MARK]
CN3 Pin # 3 6 9 12 15
0V
+0.5 to 1.0V
3
- Measure Rack Sensor Voltages using Normal Rack
and 0.1" Shim (See VP-4.1)
- Verify Measured Voltages Match Voltages in Table
SL-145
CN3
TP1
TP2
TP3
TP4
TP5
15
1
1
2
20
19
TP