CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 7.1 Mixing Paddle Overcurrent Alarm Check
3 - 113
3
Search
Book TOC
Go Back
VP - 7.1 Mixing Paddle Overcurrent Alarm Check
Purpose:
Perform this procedure to verify Mixin
g
Paddle Overcurrent Alarm.
NOTE:[NON-CE MARK] only. [CE MARK] is not adjustable.
1. Put Sample Loader in Manual Mode.
a. Press and Hold
MAN
and
START
.
b. Turn Power
ON
, hold keys down until BEEPING stops.
2. Set
BAR Switch
to
1
.
3. Remove Tower Cover.
4. Disconnect Mixin
g
Motor In-Line connector. If no In-Line connector, unsolder at Mixin
g
Motor (E
Loader - Disconnect Motor # 1(Left), Fi
g
ure 3-60).
Special Tools
DESCRIPTION QTY PN
Mixin
g
Motor
Current Tool
1 9450121