VP - 3.3 Platen Wall Adjustment
Purpose:
Perform this procedure to adjust Platen Walls, Rack Guides, and Beam Block.
1. Determine if Platen is welded or non-welded. Non-welded Platen has 14 wall screws on top of
Platen.
2. If Platen is non-welded,
o to
Step 3
. If Platen is welded,
o to
Step 4
.
3. Ali
n Non-Welded Platen:
a. Loosen Platen Wall screws (1/2 turn) at ri
ht end of Platen.
b. Insert Rack Guide Ali
nment Tool at Ri
ht end (1) of Platen. Ensure Walls are ti
ht a
ainst
Tool then ti
hten screws.
Special Tools
DESCRIPTION QTY PN
Rack Guide
Ali
nment Tool
1 9420084