VERIFICATION PROCEDURES
VP - 3.6 X-Axis Solenoid to Guide Rod Check
3 - 31
3
Search
Book TOC
Go Back
VP - 3.6 X-Axis Solenoid to Guide Rod Check
Purpose:
Perform this procedure to verify that the X-Axis Solenoid Slide is square (perpendicular)
to the X-Axis Guide Rods, Fi
ure 3-13. Required for
[NON-CE Mark]
Loaders only.
1. Locate X-Axis Home Sensor Fla
below Platen, attached to X-Axis Solenoid Slide (Fi
ure 3-12).
2. Verify color of X-Axis Sensor Fla
(Fi
ure 3-12).
GRAY [CE MARK]
- No adjustment required. Install Platen (if removed). Go to “VP - 3.8 Rack to
Laser Shield Check” (Pa
e 3-39).
BLACK [NON-CE MARK]
- Adjustment required. If Platen is removed, do not install. Go to Step
3.
3. Put Sample Loader in Manual Mode.
a. Press and Hold
MAN
and
START
.
b. Turn Power
ON
, hold keys down until BEEPING stops.
4. If Platen is installed,
o to
Step 5
, if Platen is Not installed,
o to
Step 6
.
Special Tools
DESCRIPTION QTY PN
Rack Guide
Ali
nment Tool
1 9420084
Feeler Gau
e Set 1 1-43627-01