VERIFICATION PROCEDURES
VP - 6.5 Mixing Limit Sensor Check/Adjustment
3 - 103
3
Search
Book TOC
Go Back
VP - 6.5 Mixing Limit Sensor Check/Adjustment
Purpose:
Perform this procedure to verify and Adjust Mix Assembly Limit Sensor.
E Loader
[CE MARK]
1. Put Sample Loader in Manual Mode.
a. Press and Hold
MAN
and
START
.
b. Turn Power
ON
, hold keys down until BEEPING stops.
2. Clear all racks from under mixin
g
paddle.
3. Press (Z
) key down until mixin
g
paddle stops.
4. Check that distance between platen to knurled ed
g
e of mixin
g
paddle is 3.27 in. (83mm) (Fi
g
ure
3-54).
5. If check fails, adjust Limit Sensor.
Special Tools
DESCRIPTION QTY PN
Rack Guide
Ali
g
nment Tool
1 9420084
Mixin
g
Head
Ali
g
nment Tool
1 9420082