CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 3.7 X-Axis Solenoid to Guide Rod Adjustment
3 - 36
3
Search
Book TOC
Go Back
VP - 3.7 X-Axis Solenoid to Guide Rod Adjustment
Purpose:
Perform this procedure to adjust the X-Axis Soleniod Slide so that it is perpendicular to
the X-Axis Guide Rods, Fi
ure 3-16.
NOTE:This procedure is not required for [CE MARK] Sample Loaders (Figure 3-12, Page-3-32).
1. Perform Check procedure “VP - 3.6 X-Axis Solenoid to Guide Rod Check” (Pa
e 3-31).
2. If installed, remove Platen.
3. Put Sample Loader in Manual Mode.
a. Press and Hold
MAN
and
START
.
b. Turn Power
ON
, hold keys down until BEEPING stops.
4. Remove Solenoid Umbrellas If present (TSB 97- 010, 95 - 051).
5. Place Rack Guide Ali
nment Tool on Guide Rods a
ainst Ri
ht Bracket (Fi
ure 3-16).
6. Press (
←
X
→
) keys to move X-Axis Solenoids to ed
e of tool.
CAUTION:Only use (
←
X
→)
keys
to move Solenoid slide. Moving the Slide by Hand will cause mis-
alignment.
Special Tools
DESCRIPTION QTY PN
Rack Guide
Ali
nment Tool
1 9420084
Feeler Gau
e Set 1 1-43627-01