3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING
IP - 4 Incomplete Aspiration, 2 of 2
1 - 18
1
Search
Book TOC
Go Back
IP - 4 Incomplete Aspiration, 2 of 2
Inspect and Tighten the Following Tubing Connections:
ANALYZER
- Peristaltic Pump
- Sample Valve
- 3500: Perform TSB 97-21
- Sol 96
SAMPLE LOADER
- Sample Tubing connecting Loader to Analyzer
- Vent line to Vent Needle
- Perform Cleaning Procedure (VP-12.1, Step 2)
No
If problem persists,See CELL-DYN 3000/
3500 System Troubleshooting Guide,
IP-4, Incomplete Aspiration
2
1