28 Oct
1997 20:45
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING
1.2 ERROR MESSAGES
1 - 3
1
Search
Book TOC
Go Back
1.2 ERROR MESSAGES
Table 1 - 1 Error Messages
MESSAGE
CORREC-
TIVE
ACTION
Alarms Fault Messa
eNone
Alarm 1 (v) Last messa
e sent to Analyzer was acknowled
ed ne
atively
(improper format or improper information.
IP - 21
Alarm 1 (w) Time-out - CELL-DYN did not Acknowled
e last messa
e from
Sample Loader.
IP - 21
Alarm 1 (x) Analyzer sent F command to Sample Loader, and Sample Loader
responded.
None
Alarm 1 (y) Time-out durin
aspiration. Sample Loader and Analyzer maybe
out of Sync.
IP - 21
Alarm 1 (z) Time-out durin
mixin
. Sample Loader and Analyzer maybe out of
Sync.
IP - 21
Alarm 2 Mixin
Failure IP - 22
Alarm 3 Not Used None
Alarm 4 (a) Z Ventin
/Samplin
Needle’s Ori
in and Limit Sensors BOTH ON
at Power-On.
IP - 7