VERIFICATION PROCEDURES
VP - 7.2 Mixing Paddle Overcurrent Alarm Adjustment
3 - 116
3
Search
Book TOC
Go Back
VP - 7.2 Mixing Paddle Overcurrent Alarm Adjustment
Purpose:
Perform this procedure to adjust Mixin
g
Paddle Overcurrent Alarm.
NOTE:[EUROPE] Perform this adjustment on CE MARK and NON-CE MARK loaders
[REST OF WORLD] Perform this adjustment on NON-CE MARK Loaders only.
1. Perform Check procedure “VP - 7.1 Mixin
g
Paddle Overcurrent Alarm Check” (Pa
g
e 3-113).
2. Set up Mixin
g
Paddle (“VP - 7.1 Mixin
g
Paddle Overcurrent Alarm Check” (Pa
g
e 3-113)
Steps
1-6
.
3. Adjust Overcurrent Alarm as follows:
a.
[CE MARK] Europe Only
Ensure TP 15 is cut (Fi
g
ure 3-61).
b. Adjust Potentiometer fully counter clockwise (Fi
g
ure 3-62, Pa
g
e-3-118).
[CE MARK]
Europe Only
- R64
[NON-CE MARK]
- VR10
c. While pressin
g
R
” key, adjust Potentiometer clockwise until Loader just be
g
ins to beep.
d. If F Loader, disconnect Resistor Tool and reconnect Motor. Procedure is complete. If E Loader
Go to
Step e
.
Special Tools
DESCRIPTION QTY PN
Mixin
g
Motor
Current Tool
1 9450121