CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 7.1 Mixing Paddle Overcurrent Alarm Check
3 - 115
3
Search
Book TOC
Go Back
6. Press R on keyboard to turn on mixin
g
assembly.
7. Verify Loader beeps.
For F or G Loader
g
o to
Step 11
.
• For E Loader
g
o to
Step 9
.
8. If check fails,
g
o to
Step 10
.
9. Perform the followin
g
procedure (E Loader only):
a. Reconnect Motor # 1 and connect Mixin
g
Motor Current Tool to Mixin
g
Motor #2.
b. Press
(R)
key.
- If Loader beeps,
g
o to
Step 11.
- If check fails,
g
o to
Step 10
.
10.Perform “VP - 7.2 Mixin
g
Paddle Overcurrent Alarm Adjustment” (Pa
g
e 3-116).
11.Connect motor.
E Loader Only
- Check Connection Match up.
a. Mixin
g
Motor # 1 should be connected to Blue and Purple wires.
b. Mixin
g
Motor # 2 should be connected to Yellow and Oran
g
e wires.
c. Set
BAR Switch
to
0
.
d. Press and hold (
R)
key. Only Mixin
g
Motor #1 (left) rotates.
--- END ---