3.1 OVERVIEW
This chapter contains:
Interrelationship Matrix
Adjustment and Ali
g
nment Procedures (called Verification Procedures) for the Sample Loader
(Type E, F and G)
The Interrelationship Matrix identifies the Ali
g
nments and Adjustments (VPs) required when various
parts are replaced. For example, if the Laser Shield is replaced, VP - 3.8, VP - 3.12, and VP - 11.1,
should be performed in the order indicated by the number in the VP column.
Verification Procedures ensure somethin
g
is ali
g
ned, adjusted, and workin
g
the way it should.
Usually it is checked first, then adjusted or ali
g
ned if the check is incorrect or questionable.