3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING
IP - 11 Alarm 4 (e and j) Time-out Moving X-Axis, 6 of 8
1 - 38
1
Search
Book TOC
Go Back
IP - 11 Alarm 4 (e and j) Time-out Moving X-Axis, 6 of 8
- Reseat CPU Main Bd. CN9
- BAR (1/0) Switch = 1
- Check Solenoid 1 & 2 Operation
(1 SOL 2) Keys Several Times
- Tighten Screws (Top of Solenoid)
- Ensure Solenoids are Flush or
Slightly Below Platen Surface
X-Axis
Stops Properly at
Both Left & Right
Sensors
?
X-Axis Home
Sensor at CPU Main
Bd. or Limit Sensor =
+5.0V±0.2 Blocked,
0.0V Unblocked
(See Figure)
?
Replace CPU Main Bd.
(PL-8.1; See
Interrelationship Matrix
- Pg. 3-4)
Replace
Defective Sensor
(PL-3.11; See
Interrelationship
Matrix - Pg. 3-4)
C
No
No
Yes
Yes
C1
SL-129C
TP
[NON- CE MARK]
Pin 7 X-Axis Limit
Pin 3 X-Axis Home
Pin 3
X-Axis Home
CN4
[CE MARK]
Pin 15
X-Axis Limit