CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING
IP - 15 Alarm 4 (i) Time-out Moving X-Axis, 1 of 4
1 - 56
1
Search
Book TOC
Go Back
IP - 15 Alarm 4 (i) Time-out Moving X-Axis, 1 of 4
X-Axis Moves
Properly in Both
Directions (Starts at
Slow Speed Then Goes
Full Speed)
?
1
- Power Sample Loader OFF Without Removing
Platen
- Clean and Lube Rails
- Check Obstuctions Under Platen, eg. Umbrellas,
Metallic Tape, ect.
- Enter Manual Mode
- Move X-Axis Left and Right using
X
Keys
Yes
2
No
START
X-Limit
X-Axis Sensors
X-Slowdown
X-Home
SL-123B

Table of Contents