EasyManua.ls Logo

Abbott CELL-DYN 3000 - VP3.1 Platen Pre Check

Abbott CELL-DYN 3000
296 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 3.1 Platen Pre Check
3 - 18
3
Search
Book TOC
Go Back
VP - 3.1 Platen Pre Check
Purpose:
Perform this procedure to ensure all hardware is secure, keyboard settin
g
s are correct,
and needle is home before performin
g
Platen Checks / Adjustments.
1. Remove covers.
2. Put Sample Loader in Manual Mode.
a. Press and Hold
MAN
and
START
.
b. Turn Power
ON
, hold keys down until BEEPING stops.
3. Set
BAR (0/1) Switch
to
1
.
4. Set PRE rotary switch (if used) to 0 (painted dot faces 0).
5. Press (
N
) to ensure Needle is Home (X and Y axis will not move when Needle is not home).
6. Ti
g
hten all Platen, Pusher Arm, Sensor, and Laser Shield hardware (Fi
g
ure 3-5).
7. Ensure Tower is secure and strai
g
ht (side to side and front to back). If Tower cannot be
strai
g
htened, call Tech Support.
8. Go to “VP - 3.2 Platen Wall Check” (Pa
g
e 3-20).
Special Tools
DESCRIPTION QTY PN
None

Table of Contents

Other manuals for Abbott CELL-DYN 3000