VERIFICATION PROCEDURES
VP - 6.4 Mixing Contact Sensor Check/Adjustment
3 - 91
3
Search
Book TOC
Go Back
VP - 6.4 Mixing Contact Sensor Check/Adjustment
Purpose:
Perform this procedure to verify and adjust the Tube Contact Sensor Position and
Sensitivity (Fi
g
ure 3-45) .
1. Remove all Racks from Sample Loader.
2. Put Sample Loader in Manual Mode.
a. Press and Hold
MAN
and
START
.
b. Turn Power
ON
, hold keys down until BEEPING stops.
3. Set
BAR (1/0) Switch
to
1
.
4. Ensure Mix Paddle(s) move Up/Down freely. Push up with fin
g
er, Paddle should not stick.
Special Tools
DESCRIPTION QTY PN
Feeler Gau
g
e Set 1 1-43627-01
Mixin
g
Head
Ali
g
nment Tool
1 9420082
Rack Guide
Ali
g
nment Tool
1 9420084