Commonly Used Procedures 6
Agilent Nano Indenter G200 User’s Guide 6-35
Calibrate Tip (rare)
Figure 6-49Calibrate commands in Configure Tip Parameters dialog
box
When calibrating, be sure to use inputs relevant to the head used to
acquire the data.
Figure 6-50 shows the Tip Area and Frame Stiffness Calibration dialog
box used for calibration inputs using the Area Function Options tab.