Commonly Used Procedures 6
Agilent Nano Indenter G200 User’s Guide 6-64
16 Choose clean location on reference material and set array size of at
least ten indentation tests.
17 Click the Next Step button.
18 Click Run Batch.
Periodically, NanoSuite will display the message shown in Figure 6-89.
This is normal because the indenter has not yet touched the surface of
the sample.
Figure 6-89NanoSuite message
Allow testing to complete.
19 When testing is complete, go to the Review page.
20 Right-click anywhere in the Legend pane and select Select All Tests
from the resulting pop-up menu.
21 Delete any obviously aberrant tests, such as the aberrant test shown
in
Figure 6-81 on page 6-57:
a Identify the test number by referring to the legend next to the
graph of results.
b Right-click on the number of the aberrant test in the Legend pane.
c Select Delete Test from the resulting pop-up menu.
d Click Ye s in the Test Deletion confirmation box
e Right-click anywhere in the Legend pane and select Select All
Tests from the resulting pop-up menu to populate the graph.
22 Access the frame stiffness and area function information by choosing
Tip > Configure Tip as shown in
Figure 6-90.
Figure 6-90Select Configure Tip from Tip menu