NanoSuite Explorer D
Agilent Nano Indenter G200 User’s Guide D-58
Figure D-67Selecting a GoTo test segment
46 Note the properties in the Settings pane.
47 Right click on this test segment and select Insert before from the
resulting pop-up menu.
48 In the Insert Test Segments dialog box, select a GoTo type test
segment.
49 Click OK.
50 In the Settings pane, set the following properties:
a Direction = Loading
b Rate Value = Ramp Loading Rate
c Enable Stopping Conditions = selected
d Stop Channel = Load On Sample
e Comparison = Exceeds
f Stop Value = Ramp Force Limit
g Run-Time Message = Initial ramp load
51 On the left-hand side of the screen, click on the Configuration tab.
52 Click on the Miscellaneous Configuration item.
53 In the Settings pane, edit the Method Description to the text shown
in
Figure D-68 on page D-59.