Express Test I
Agilent Nano Indenter G200 User’s Guide H-6
7 Click the Run button.
Once testing is complete, the software will prompt the user to save the
file.
8 Input a file name and click Save.
9 To view the results, click the Review tab.
The 2D data is available by selecting the Channels tab as shown in
Figure H-4 on page H-6
Figure H-4Examples of Results from the Channels tab
The available results are:
Points Excluded: Points rendered invalid and excluded from reported
results by the Filter Graph function (Help menu provides further
information on this subject).
Disp Into Surface (nm): Peak indentation displacement, averaged over
all valid indents in the test array.
Modulus (GPa): Young's modulus, averaged over all valid indents in
the test array.
Modulus SD (GPa): Standard deviation of valid Young's modulus
measurements for the test array.
Hardness (GPa): Hardness, averaged over all valid indents in the test
array.
Hardness SD (GPa): Standard deviation of valid hardness
measurements for the test array.
Stiffness (N/m): Contact stiffness, averaged over all valid indents in the
test array. If the contact stiffness is more than 10
% of the frame
stiffness, best results may require the determination of frame stiffness