AS Autosampler Operator’s Manual
102 Doc. 065051-03 1/08
If the AS never reaches a status that allows opening the door, check for one of
the following:
If the AS is operating under Chromeleon or Chromeleon Xpress control,
check the run length specified for the program. The run length may need
to be increased to allow a longer wait period between injections.
If the AS is operating under front panel control, verify that a cycle time
was specified for each method in the schedule and that the cycle time is
long enough to provide a wait period between injections.
3.10 Routine Maintenance
This section describes routine maintenance procedures that the user can perform.
Any other maintenance procedures must be performed by Dionex personnel.
3.10.1 Daily
Check for air bubbles in the syringe(s) and tubing and remove them if
they appear (see Section B.3.14
).
Check the volume of liquid in the flush bottle (and any reagent bottles
that are installed) and refill when needed. After refilling the flush
reservoir, manually flush the inject port (see Section 3.10.4
).
Check the volume of liquid in the waste container and empty when
needed.
Neutralize acidic and caustic wastes before disposal. Dispose of all
wastes in accordance with local regulations.
Neutralisez les déchets acides ou caustiques avant de les jeter. Jetez
les déchets aux règlements locaux.
Neutralisieren Sie säurehaltige und ätzende Abfälle vor ihrer
Entsorgung. Entsorgen Sie alle Abfälle entsprechend den lokalen
Bestimmungen.
For correct drainage, make sure the end of the waste line is not
submerged in waste liquid.