5•Service
Doc. 065051-03 1/08 131
5.2.2 Resetting a Usage Counter on the Usage Log Screen
1. On the MENU of SCREENS, press 8, 6, and 1 to go to the USAGE LOG
screen (see Figure 5-2
).
The monitored parts do not all fit on one screen. If necessary, press
the down arrow to display the usage counter to be reset (see
Figure 5-3
).
2. Position the cursor in the
RESET field of the part that was replaced or
serviced and press
Enter.
The number is reset to zero.
Figure 5-2. Usage Log Screen
Figure 5-3. Usage Log Screen (Additional Parameters)
Help Prompt
USAGE LOG
MODULE ON:
BACKLIGHT:
SAMP SYRINGE USE:
PREP SYRINGE USE:
INSERTS INJ PORT:
nnnnnn hours
nnnnnn hours
nnnnnn injects
nnnnnn cycles
nnnnnn cycles v
RESET
*
Help Prompt
USAGE LOG
MODULE ON:
BACKLIGHT:
INJECT VALVE:
DIVERTER VALVE:
XYZ NEEDLE ASSEM:
nnnnnn hours
nnnnnn hours
nnnnnn cycles
↑
nnnnnn cycles
nnnnnn cycles
RESET
*