3 Operation and Maintenance
Doc. 065051-03 1/08 57
3.4.2 Well Plate Tray Sample Positions
NOTE Chromeleon or Chromeleon Xpress is required
to sample from a well plate.
Sample positions in the well plates are specified with a two letter and one
number identifier (for example, LB8):
The first letter (either L or R), specifies the left or right well plate (see
Figure 3-9
).
The second letter (A through H) and the number (1 through 12)
specify the well position in the well plate. Well letters and numbers
are printed on well plates and well plate covers.
For example, well plate position LA1 identifies the well located at
position A1, in the well plate installed on the left side of the AS well plate
tray. When the well plate is installed in the tray, position LA1 is in the
lower-left corner of the tray (see Figure 3-9
).
NOTE To specify the position of a 1.5 mL vial installed in the
well plate tray, precede the vial position number with a
V (V1, V2,V10) (see Figure 3-9
).
Figure 3-9. Well Plate Tray: Well Plate and Vial Positions
12 3456789101112
A
B
C
D
E
F
G
H
RA1
RH12
12 3456789101112
A
B
C
D
E
F
G
H
LA1
LH12
L
A1
R
A1
12345678910
V1
V10
Position
LA1