3 • Operation and Maintenance
Doc. 065051-03 1/08 55
3.3 Default Operating Parameters
The Table 3-1 lists the factory-set values for AS operating parameters:
Operating Screen Parameter Default Value
Method Sample Prep Step 1 wait step only
Method Setup Cycle Time
Sample Needle Height
Sample Temperature
Wait for Temp Stable
Off
2 mm
20 °C
No
Method Timed
Events
Time=Init
Inject Valve
TTL/Relays
Load (for simultaneous or sequential
injection, both valves at Load position)
All off (0)
Time=0.00
Inject Valve
TTL/Relays
Inject (for simultaneous or sequential
injection, both valves at Inject position)
All off (0)
Schedule Missing Vial Action
Vial # Start/End
Injections per Vial
Injection Volume
Method Number
Stop
1/1
1
25.0 μL; 1 mL for simultaneous injection
1
System Parameters Wait
Syringe Speeds
Cut Volume
Sample Mode
Inject Port Volume
Air Bubble Volume
(not user changeable)
Enabled
Refer to Section 3.7
for Chromeleon control
syringe speeds. Refer to Section D.7
for
front panel control syringe speeds.
10 μL (see Section 3.6.3
for details)
Normal
varies (each inject port is individually
calibrated)
8 μL if the sample syringe is <5 mL;
60 μL if a 5 or 10 mL syringe is installed
Flush/Prime Flush Volume
Flush Reservoir
Flush Syringe
Prime Volume
Prime Reservoir
Prime Syringe
250 μL
Flush (always)
Sample (always)
400 μL
Flush
Sample
Table 3-1. Default Operating Parameters