A • Specifications
Doc. 065051-03 1/08 177
A.2 Sample Temperature Control (Optional)
NOTE The temperature inside the autosampler compartment
may exceed ambient by up to 2 °C. If samples are
temperature sensitive, Dionex recommends installing
the sample temperature control option/
Injection Loop Size 25 μL standard; other sizes are available
Concentrator
Column
Dionex ultra-low pressure; maximum injection pressure is 690 kPa
(100 psi)
Maximum Pressure
Delivery
690 kPa (100 psi)
Temperature
Control Range
4
to 60 °C in 1 °C increments at 24 °C (75 °F) ambient, stable to
±0.2 °C
Operating
Humidity
5 to 90% relative humidity, noncondensing
Cooling capacity is slightly reduced at high humidity
Dimensions
(Sample Temp. +
Autosampler)
65 cm high x 30 cm wide x 46 cm deep (2.75 in x 12 in x 19 in)
Weight
(Sample Temp. +
Autosampler)
<38 kg (<81 lb)