B • Installation
Doc. 065051-03 1/08 207
B.3.10 Turning On the Power
Press the power switch in the lower left corner of the front door to turn on
the
AS power. When the power is turned on, a series of screens are
displayed and the autosampler performs a power-up sequence.
• The COPYRIGHT NOTICE screen is displayed briefly, and then the
POWER-UP screen appears.
• While the POWER-UP screen is displayed, the AS performs
initialization procedures and a series of diagnostic tests. If one of the
tests fails, testing stops and the
DIAGNOSTIC TEST screen is
displayed. See Section C.9.2
if this occurs.
• When the initialization procedures are complete and the AS has
passed the diagnostics, the
INSTALLED OPTIONS screen appears. After
a few seconds (or if you press a button), the
MAIN screen is displayed.
The power-up sequence is then complete.