Doc. 065051-03 1/08 169
5.15 Replacing a Sample or Prep Syringe Pump Assembly
The following syringe pump assemblies are available: Sample syringe
(P/N 062382), dual sample syringe (P/N 062379), and sample prep syringe
(P/N 062365)
1. Turn off the AS power.
2. Disconnect the liquid lines from the existing syringe valve.
3. Unplug the syringe cable from the rear panel connector.
4. Lift up the syringe pump assembly and remove it from the organizer.
5. Set the new syringe pump assembly into the organizer.
6. Connect the liquid lines.
7. Plug in the cable.
8. Turn on the AS power.