Doc. 065051-03 1/08
5•Service
This chapter describes AS Autosampler service and repair procedures that the user
can perform. All procedures not included here, including electronics-related repair
procedures, must be performed by Dionex personnel. For assistance, contact
Dionex Technical Support. In the U.S., call 1-800-346-6390. Outside the U.S.,
call the nearest Dionex office.
Before replacing any parts, refer to the troubleshooting information in Chapter 4
to isolate the cause of the problem.
The CPU card contains a lithium battery. If the CPU card is
replaced, dispose of the used battery properly. Contact your local
government for disposal or recycling practices in your area.
5.1 Replacement Tubing and Fittings
The following tables list the tubing, fittings, and tubing assemblies used to plumb
the AS autosampler compartment.
NOTE The AS is capable of delivering at a maximum pressure
of 690 kPa (100 psi). Refer to Section 3.8
to select the
appropriate syringe speed and tubing sizes to keep the
total backpressure below 690 kPa (100 psi).
5.1.1 Syringe and Reservoir Connections
Substituting non-Dionex parts may impair AS performance, thereby
voiding the product warranty. Refer to the warranty statement in the
Dionex Terms and Conditions for more information.
Part Number Description Used For
052112
048949
052267
PFA tubing: 1.55-mm (0.062-in) ID
1/8-in flangeless ferrule fitting
1/4-28 flangeless nut
Connecting the flush reservoir
to the sample syringe valve