AS Autosampler Operator’s Manual
176 Doc. 065051-03 1/08
A.1.3 Valves (Optional)
A.1.4 Injection
Injection One or two, two-position, six-port, electrically-activated Rheodyne
valves, PEEK
Diverter One two-position, six-port, electrically-activated Rheodyne valve,
PEEK
Injections per Vial 1 to 99
Vial Sizes Available 0.3 mL polymer vials
1.5 mL glass vials
10 mL PolyVials
Well Plate Sizes
Available
Deep-well plate, 96 wells, 2.0 mL
Micro-well plate, 96 conical (V-shaped) wells, 0.45 mL
Micro-well plate, 96 round (U-shaped) wells, 0.5 mL
Varia ble Volu me
Range
If the sample syringe is 1,000 μL or less:
1.0 to 99.9 μL in 0.1 μL increments; 100 to 1,000 μL in
1 μL increments (normal or sequential mode)
100 to 1,000 μL in 1 μL increments (concentrate or sequential
concentrate mode)
If the sample syringe is 5,000 μL:
1.0 to 99.9 μL in 0.1 μL increments; 100 to 4,700 μL in
1 μL increments (normal or sequential mode)
100 to 4,700 μL in 1 μL increments (concentrate or sequential
concentrate mode)
1,000 to 4,700 μL in 1 μL increments (simultaneous mode)
If the sample syringe is 10,000 μL:
1.0 to 99.9 μL in 0.1 μL increments; 100 to 8,000 μL in 1 μL
increments (normal or sequential mode)
100 to 8,000 μL in 1 μL increments (concentrate or sequential
concentrate mode)
1,000 to 8,000 μL in 1 μL increments (simultaneous mode)