4.3 Condensation on Well Plate Covers
Condensation can form on the well plate covers of the micro-well plates if the tray
temperature set point is at 4 °C and the ambient temperature and relative humidity
in the laboratory are above 25 °C and 60% humidity.
To prevent condensation, either ensure that the laboratory temperature remains
below 25 °C and the humidity below 60%, or increase the tray temperature set
point.