C • Display Screens
Doc. 065051-03 1/08 257
C.7 Flush/Prime Screen
Use the FLUSH/PRIME screen to:
• Set the parameters for and then initiate a flush cycle that flushes the inject port
and washes the outside of the needle.
• Set the parameters for and then initiate a prime cycle that primes the lines to
the reservoir(s) and the sampling needle.
NOTE To make sure the backpressure during flushing remains
below 690 kPa (100 psi), select a flush syringe speed that
is appropriate for the syringe size and tubing volume.
See Section 3.8
for details.
If the AS is controlled by Chromeleon or Chromeleon
Xpress, set the syringe speed on the Chromeleon or
Chromeleon Xpress Control panel (see Section 3.7
). If
the AS is always controlled from the front panel, select
the syringe speed on the
SYSTEM PARAMETERS screen
(see Section C.6.5
).
Screen Field Description
DOOR INTERLOCK
BYPASS
Press Select Δ or Select ∇ to select YES (bypass the door
lock) or
NO (do not bypass the door lock) and then press Enter.
The default setting is
NO.
After selecting
YES and pressing Enter, the door lock alarm is
disabled for 15 minutes. If you open the autosampler
compartment door during this time, the sampling arm continues
its current operation. However, if a schedule is running, opening
the door ends the schedule. To extend the bypass for another
15 minutes, press Enter again.
When the 15 minutes is up, the
DOOR INTERLOCK BYPASS
setting automatically reverts to
NO. When NO is selected,
opening the autosampler compartment door stops the sampling
arm operation and ends the currently running schedule.