5•Service
Doc. 065051-03 1/08 153
3. Under Wellness, click the Needle align button. The Needle Alignment panel
appears (see Figure 5-25
).
4. To complete the alignment procedure, follow the instructions on the panel and
refer to the following notes:
Step 1: For a better view of the needle and inject port during the alignment
procedure, you can open the autosampler door after clicking Inject port
align.
Figure 5-25. Chromeleon Needle Alignment Panel
Be aware of unexpected sampling arm movement when operating with
the door open.
Soyez conscient de la possibilité de mouvements imprévus du bras
d'échantillon lorsqu'il est utilisé avec la porte ouverte.