C Display Screens
Doc. 065051-03 1/08 239
FLUSH
Flushes the needle with a specified volume, from 1 to 999 μL
(default=250 μL). The sample syringe and flush reservoir are
always used for flushing. If the volume specified is greater than
the syringe volume, the syringe takes multiple strokes to achieve
the volume.
This operation is independent from the flush that takes place
before each injection in a schedule (see Section 2.7.6
).
DISPENSE
(sample prep
option only)
Moves a specified volume of reagent from a reservoir to a vial
using the prep syringe.
DISPENSE is available only if the
sample preparation option is installed.
μ
L: Dispense volume: from 100 to 9,999 μL
(default=1,000 μL).
source: Source reservoir (the reservoir from which reagent is
dispensed):
RES A, RES B, RES C, or RES D
(default=RES A).
to #: Destination vial position (the vial to receive
the dispensed volume): (default=
SMP).
DILUTE
(sample prep
option only)
Dilutes a volume of concentrate with a specified volume of
diluent.
DILUTE is available only if the sample preparation
option is installed.
μ
L: Concentrate volume: from 1.0 to 99.9 μL
or 100 to 1,000 μL (default=10 μL).
#: Source vial: The sample syringe picks up the
concentrate from the source vial and dispenses it to the
destination vial (default vial=
S+1).
μ
L: Diluent volume: from 10.0 to 99.9 μL
or 100 to 9,999 μL (default=10 μL) (default=990 μL).
source: Diluent reservoir (the reservoir from which diluent is
dispensed):
RES A, RES B, RES C, or RES D
(default=
RES A).
to # : Destination vial position (the vial that the concentrate
and diluent volumes are delivered to): (default
vial=
SMP).
Prep Operation Description