C Display Screens
Doc. 065051-03 1/08 241
Notes on Reagent Volume Ranges
If the source is a reservoir, the allowed range is 100 to 50,000 μL. The
default is 2,000
μL if the syringe volume is less than 5,000 μL, and
10,000
μL if the syringe volume is 5,000 μL or 10,000 μL.
If the source is a vial, the range depends on the volume of the sample
syringe:
NOTE The REAGENT FLUSH and REAGENT PRIME operations
are performed after the autosampler loads the sample
onto the concentrator column. See Section 3.6.10
for
details about concentrator injections.
Sample Syringe
Volume (µL)
Reagent Flush
Range (µL)
Default (µL)
100 to 1,000 100 to 1,000 1,000 (any size vial)
5,000 100 to 4,700 1,000 (1.5 mL vial)
4,000 (10 mL vial)
10,000 100 to 8,000 1,000 (1.5 mL vial)
8,000 (10 mL vial)