AS Autosampler Operator’s Manual
302 Doc. 065051-03 1/08
LINE
After you enter the START and END vial numbers, the LINE field displays
the appropriate schedule line numbers. You cannot edit the
LINE field.
START and END
Enter the starting and ending vials for a range of vials to be sampled.
A range is a group of consecutively numbered vials that will be
sampled using the same schedule parameter values (injections per
vial, injection volume, and method number). The starting vial can be
a higher number than the ending vial.
For a single vial, enter the same number for START and END.
NOTE To sample from a well plate, the AS must be under
Chromeleon or Chromeleon Xpress control.
INJ/VIAL
Enter the number of injections (0 to 99) to perform on each vial.
If you enter 0, only the sample prep phase of the method is run.
If you enter multiple injections per vial, the sample prep phase of the
method is performed once only (before the first injection). The
method setup and timed events are performed for each injection.
INJ VOL
Enter the volume of sample to deliver for each injection. The range
allowed depends on the sampling mode and the size of the sample
syringe:
Tray Type Tray Capacity Valid Vial Numbers
1.5 mL, aluminum 100 vials (1.5 or 0.3 mL) 1 through 100
10 mL, plastic 49 PolyVials 1 through 49
Sampling Mode Sample Syringe
Size
Inject Volume
Ranges
Default
Normal Any 1.0 1,000 μL25.0μL
Simultaneous 5 mL
10 mL
1,0004,700 μL
1,0008,000 μL
1,000 μL
1,000 μL
Concentrator 1001,000 μL
5mL
10 mL
1001,000 μL
1004,700 μL
1008,000 μL
1,000 μL
4,000 μL
8,000 μL