Frequency The default system frequency is 50/60 Hz. Use the
Setup display (Figure 3.3 on page 3-5) to change
the default frequency. Use the spinner arrows in
the Frequency column to select the AC harmonic
or to select a DC ‘+’ or a DC ‘–’ range.
Range The range setting determines the maximum value
for the amplitude of a particular source. For
maximum compliance voltage, use the lowest
current range that can produce the desired test
current. See Appendix G ”F6150 Specifications”
for more details about range settings.
The last two columns in the source table contain On and Enable
buttons for each source.
On Click the On button to activate a source. The
button turns red, and the System Output button
blinks red. Click the On button again to turn
a source off, the button turns gray. The System
Output button stops blinking when all the sources
have been turned off.
Enable Click the Enable button to place selected sources
in standby status. The Enable button for each
source to be activated turns green. When System
Output is selected, all of the enabled sources turn
on. The System Output button turns red, and the
On buttons for the individual sources stay gray.
Clicking System Output again turns the enabled
sources off.