IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
8-6 Section 8-2 - Covers - Replacement Procedures
PRELIMINARY
8-2-1-1 Preparations
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in Power Shut Down on page 4 - 7.
8-2-1-2 Order of Cover Removal
Remove the system covers in the following order, as applicable:
Air Inlet (Left Side) Cover, as described on page 8-7.
Right Side Cover, as described on page 8-10.
Either: DVD Cover, as described on page 8-11
Or: DVD and Printer Cover, as described on page 8-14
Front Cover, as described on page 8-16.
Right Rear Cover, as described on page 8-20.
Left Rear Cover, as described on page 8-23.
AC Distribution Box Cover, as described on page 8-29.
Note: The Vivid™ S60N/Vivid™ S70N ultrasound scanner has additional covers on the control
console components (Monitor and Touch Screen). Instructions for removal of these covers are
described in the relevant replacement procedure sections.
Note: For removal of system accessories, refer to the following procedures:
OPIO Basket Replacement Procedure on page 8 - 26
Rear Folder Box Replacement Procedure on page 8 - 27
8-2-1-3 Preparation for Cover Installation
Replacement covers for the Vivid™ S60N/Vivid™ S70N ultrasound scanner are supplied with the
required securing screws and ball stud/ball stud receptacles. Before installing a replacement cover, it is
necessary to fit these in the appropriate positions. Refer to the illustrations provided in the specific
Cover Replacement procedure - for example, AC Distribution Box Cover:.
DANGER
ELECTRICAL HAZARDS EXIST AT SEVERAL POINTS IN THE SYSTEM.
FAMILIARIZE YOURSELF WITH ALL HAZARDOUS VOLTAGES AND
HIGH CURRENT LEVELS BEFORE REMOVING ANY OF THE COVERS.
DANGER
DO NOT WEAR THE ESD WRIST BAND STRAP WHEN REMOVING
PARTS FROM THE POWER SUPPLY UNIT. BEFORE REMOVING ANY
PART OF THE POWER UNIT, TURN THE POWER OFF AND
DISCONNECT THE POWER CORD.
CAUTION
BEFORE REMOVING CIRCUIT BOARDS, TURN THE POWER OFF AND WEAR THE ESD WRIST
BAND STRAP.