IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
8-70 Section 8-4 - Cables - Replacement Procedures
PRELIMINARY
11.)Remove the cable securing clip and cut the tie-wraps - see Figure 8-88 below.
12.)Thread the cables through the OPIO platform aperture - see Figure 8-89 below.
Figure 8-88 Remove Securing Clip and Tie-Wraps on OPIO Platform
Figure 8-89 OPIO Platform Showing Cables Threaded Through Aperture