IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-177
PRELIMINARY
6.
Remove the Software Installation Media from the USB
port, then click Yes to continue.
Important:
While the screen remains blank, do NOT attempt
to use the system or interrupt the installation
process!!
The system performs a reboot.
On completion, the Windows icon and message "Starting Windows" is
displayed on the Touch Screen.
A series of setup messages are displayed on the Vivid™ S60N/Vivid™ S70N
ultrasound scanner Monitor, in the following sequence:
"Setup is starting services"
"Setup is installing devices"
"Setup is applying system settings"
"Setup will continue after restarting your computer"
The system then performs automatic shutdown, then reboots.
An additional messages is now displayed on the Monitor:
"Setup is checking video performance"
On completion of the above, the Monitor LCD display will remain blank for
~30 secs
The cursor will be visible at the top-left corner of the screen:
The Welcome screen is then displayed:
Table 8-4 Windows Operating System and Vivid™ S60/Vivid™ S70 Software Installation Procedures
Step Instruction Expected Result