D
IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-183
PRELIMINARY
18.
Type the appropriate SW License key (option key).
Refer to Preparation for Software Installation
Procedures on page 8 - 174.
Check the number is entered correctly, then click OK.
The dialog box closes; the Vivid S60N/Vivid S70N Startup Screen is displayed,
followed by the Scanning screen.
19.
Press Config, then click About at the bottom of the
screen. Select the System Version tab.
The system version details are displayed.
20.
Verify the Application Software version corresponds
with the System Software version you have installed
(check against the label on the Software Installation
Media) and that it was correctly installed.
21.
Restore Patient Archive and System Configuration
as per User Manual.
22.
Install peripherals.
23.
Restore the settings as recorded under Preparation
section 8-7-1-1 on page 8-174.
Verify proper Connectivity functionality - refer to
Connectivity Setup on page 3 - 41.
Table 8-4 Windows Operating System and Vivid™ S60/Vivid™ S70 Software Installation Procedures
Step Instruction Expected Result