176
────────────────────────────────────────────────────
11.4 Command Reference
────────────────────────────────────────────────────
■ Current auto range setting
Synta
CURRent:AUTO data
dat
ON, OFF (character data)
Functio
Turns current auto range.
Erro
・In the following circumstances, a device dependent error is generated:
When this command is executed in the hold state.
If an attempt was made to execute this command while waiting for or
executing integrated measurement or demand measurement.
In the event of a system error.
・If data is set to character data other than "ON" or "OFF", an execution
error is generated.
Exampl
Transmission
":CURR:AUTO ON"
Current auto range setting is turned on.
■ Current auto range setting query
Synta
CURRent:AUTO?
Functio
Returns as "ON" or "OFF" (data) whether or not current (amperage) auto
ranging is presently enabled.
Respons
synta
If headers are ON
If headers are OFF
:CURRENT:AUTO data"
"data"
Exampl
Transmission
Response
":CURR:AUTO?"
":CURRENT:AUTO ON"
":CURR:AUTO?"
"ON"
URRent:AUTO
URRent:AUTO?