176
────────────────────────────────────────────────────
11.4 Command Reference
────────────────────────────────────────────────────
Current auto range setting
Synta
x
CURRent:AUTO data
dat
a
ON, OFF (character data)
Functio
n
Turns current auto range.
Erro
r
In the following circumstances, a device dependent error is generated:
When this command is executed in the hold state.
If an attempt was made to execute this command while waiting for or
executing integrated measurement or demand measurement.
In the event of a system error.
If data is set to character data other than "ON" or "OFF", an execution
error is generated.
Exampl
e
Transmission
":CURR:AUTO ON"
Current auto range setting is turned on.
Current auto range setting query
Synta
x
CURRent:AUTO?
Functio
n
Returns as "ON" or "OFF" (data) whether or not current (amperage) auto
ranging is presently enabled.
Respons
e
synta
x
If headers are ON
If headers are OFF
:CURRENT:AUTO data"
"data"
Exampl
e
If headers are ON
If headers are OFF
Transmission
Response
":CURR:AUTO?"
":CURRENT:AUTO ON"
":CURR:AUTO?"
"ON"
C
URRent:AUTO
C
URRent:AUTO?

Table of Contents