179
────────────────────────────────────────────────────
11.4 Command Reference
────────────────────────────────────────────────────
Demand measurement integrated output rate setting
Synta
x
DEMand:AOUT:RATE data1,data2,data3,data4
dat
a
1, 2, 3, 4, 5, 6 (numerical data in NR 1)
1: 5 V/5 kWh
2: 5 V/10 kWh
3: 5 V/50 kWh
4: 5 V/100 kWh
5: 5 V/500 kWh
6: 5 V/1 MWh
Functio
n
Sets the output rate when an integrated value was specified for the D/A
output in Demand measurement mode.
Erro
r
An execution error results if any value other than that indicated above is
set for (data).
In the following circumstances, a device dependent error is generated:
If an attempt is made to execute this command while waiting for or
during demand measurement.
In the event of a system error.
Exampl
e
Transmission
":DEM:AOUT:RATE 1"
Sets the Demand mode D/A output rate to "5 V/5 kWh".
Demand measurement integrated output rate query
Synta
x
DEMand:AOUT:RATE?
Functio
n
Returns the Demand measurement mode D/A output rate setting.
Respons
e
synta
x
If headers are ON
If headers are OFF
":DEMAND:AOUT:RATE data"
"data"
Exampl
e
If headers are ON
If headers are OFF
Transmission
Response
":DEM:AOUT:RATE?"
":DEMAND:AOUT:RATE 1"
":DEM:AOUT:RATE?"
"1"